Infant Radiant Warmer Infant Incubator (AM-100A)

INFANT RADIANT WARMER Infant Incubator (AM-100A)Based on AM-100, AM-100A is added with the side door which can be opened so that the bassinet can be pulled out.Specification:Air mode control, set temperature and air temperature are displayed separately;Natural air flow humidity;Failure alarm indication;RS-232 connector.NOTICE:If

Featured Products

INFANT RADIANT WARMER Infant Incubator (AM-100A)
Infant Radiant Warmer Infant Incubator (AM-100A)

Infant Radiant Warmer Infant Incubator (AM-100A)

Based on AM-100, AM-100A is added with the side door which can be opened so that the bassinet can be pulled out.

Specification:
Air mode control, set temperature and air temperature are displayed separately;
Natural air flow humidity;
Failure alarm indication;
RS-232 connector.

NOTICE:
If the product standard is changed, the parameters are subject to actual products.
The product's technic can be changed according to customer need under the product norm.

Technical specification:
Power supply: AC220V, 50Hz
Power input: 600VA
Air Temperature control range: 25º C~37º C
Sensor precision: ≤ 0.3º C
Changing of temperature: ≤ 0.5º C
Temperature rising time: ≤ 30min
Noise level: ≤ 55dB(A)
Inclination of bassinet: 0~5º C

Alarming:
Audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater (except for the low deviation alarm), meanwhile, the skin temperature indication window can indicate the alarm code. For the alarm 1-6, the relevant light is on, and for the alarm 7, all alarming lights are on; All alarm sound can be cancelled except for the power failure alarm and the system alarm.
Over-temp alarm: When the temperature ≤ 38º C;
Upper Deviation alarm: Air temperature control 3.0º C
Lower Deviation alarm: Air temperature control -3.0º C
Sensor alarm: When the sensor is disconnected, open circuit, short circuit or put on the wrong place;
Fan alarm: When the fan is blocked, speed ≤ 1000rpm or stop working;
Power failure alarm: When there is no power or the power cord is disconnected.
System alarm: When there is E2ROM malfunction inside of temperature controller.

Air mode control,set temperature and air temperature are displayed separately;
Natural air flow humidity;
Failure alarm indication;
RS-232 connector.
Package size: 1085mm×640mm×1120mm
Net weight: 80kg
Part NumberNames of PartsParts Pictures
02.38.007.01Controller
 
02.02.005.03SHood
 
01.04.025.01Air filter
 
86.22.004.00Wheel
 
86.22.004.00Wheel
 
06.60.002.01SI.V. pole groupware
 

For more information, please visit our website or contact us directly with your questions or inquiries. We are looking forward to receiving your correspondences in the near future.